Trace transmission, how much is the error between channels?

Add time:2024-08-12   Number of views:88  

cb4b92d6d92931cf721e78ae0125f6a.jpg

Through the TOUCH series intelligent pump assembly 4 channel NC0410 pump head experiment to verify the inter-channel error value, through the amount of liquid 1mL for example, the pump head in the inter-channel error to achieve the ultimate, the maximum inter-channel error is not more than 2% or even 0.16%.

513c8c0959e321c2144128a2c13cbec.jpg

Through the experiment, it can be concluded that the NC series pump head can solve the influence of channel error to the greatest extent, and provide a strong guarantee for the experiment with high precision!

In addition, the repeatability test of a single channel is also done, and the maximum repeatability error is less than 1%. This proves that both peristaltic pumps and pump heads have very high stability.